Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument

Fauziah Fauziah, Bualkar Abdullah

Abstract


This research in determining x-ray effective energy in linear accelerator (Linac) radiotherapy device was done using copper plates at different thicknesses by determining the Half Value Layer (HVL) for each energy. The purpose of this research is to know the amount of effective energy used in radiotherapy linear accelerator. The results showed that the energy of 6 MeV obtained 2,714 MeV of effective energy at HVL 4,500 mm and at the energy of 10 MeV obtained 4,231 MeV of effective energy at HVL 4,500 mm.

Keywords


copper plates; effective energy; half value layer; linear accelerator

Full Text:

PDF

References


E Ariga, S., Ito, S., Deji, T., Saze, K., & Nishizawa. (2012). Determination of Half Value Layers of X-ray equipment Using Computed Radiography Imaging Plates. Physics Medica, 1(1). https://doi.org/doi:10.1016/j.ejmp.2011.01.001

Fahmi, Arif, K., Firdausi, S., & Budi, W. S. (2008). Pengaruh Faktor Eksposi pada Pemerikasaan Abdomen terhadap Kualitas Radiograf dan Paparan Radiasi Menggunakan Computed Radiograph, 11(4), 109–118.

G., W., & Budiantari, C. T. (2005). Optimalisasi Aspek Keselamatan Pada Kalibrasi Pesawat Radioterapi.

Gabriel, J. (1996). Fisika Kedokteran. Denpasar-Bali: Departemen Fisika Universitas Udayana.

Nurdin, W. B. (2015). Fisika Kedokteran Nuklir, Pusat Kajian dan Peningkatan Aktivitas Instruksional LKPP. Makassar: Universitas Hasanuddin.

Palinggi, Y. (2016). Penentuan Nilai Half Value Layer (HVL) Dari Plate Aluminium Dengan Metode Variasi Ketebalan Filter Menggunakan Pesawat Linear Accelerator (LINAC). In Universitas Hasanuddin. Makassar: Universitas Hasanuddin.

Peraturan Pemerintah Republik Indonesia. Keputusan Menteri Kesehatan Republik Indonesia Nomor 1250/MENKES/SK/XIII/2009 Tentang Pedoman Kendali Mutu (Quality Control) Peralatan Radiodiagnostik, Pub. L. No. 13 (2009). Indonesia.

Sprawls, P. (1995). Physical Principles of Medical Imaging (2nd ed.). USA: Aspen Publisher.

Sumarni. (2008). Pusat Pendidikan dan Pelatihan Badan Tenaga Nuklir Nasional. In Deteksi dan Pengukuran Radiasi. Jakarta: Pusdiklat-BATAN.

Sumarni, T., & Mart, N. (2000). Energi Efektif Sinar-X Radiodiagnostik. In Puslitbang Keselamatan Radiasi dan Biomedik Nuklir. Batan.

Yuniasari, L., Kristiyanti, B., Rozali, B., & Syawaludin. (2013). Perancangan Perisai Radiasi pada Kepala Sumber untuk Pesawat Radioterapi Eksternal Menggunakan co-60 pada Posisi beam off. Pusat Rekayasa Perangkat Nuklir (PRPN), 7.

Yunita, H. D., Setiawati, E., & Anam, C. (2014). Evaluation Method Determination Half Value Layer (HVL) Using Multi Purpose Detector (MPD) Barracuda Sinar-X Mobile, 3(2), 113–118.




DOI: http://dx.doi.org/10.24042/jipfalbiruni.v7i1.2336